Back to overview- Designation of the standard used for the hardness test
- Description of the specimen
- Measured values must be specified in accordance with the specifications stipulated in the standard (Point 4.2)
- Date of test
- Temperature during the test, if outside the range 10 °C to 35 °C
- If the ratio of the indentation diameter to the diameter of the indenter is outside the range 0.24 – 0.60, then this must be specified in the test report.
- If a conversion is used, the conversion process and its underlying algorithm must be specified
- Information about the test, which is not specified in ISO 6506
- Information about any occurrences that may have had an influence on the result